Wheatstone Bridge Calculator – Unknown Resistance at Balance
Use this free Wheatstone Bridge Calculator to find the unknown resistance Rx in a balanced Wheatstone bridge. Enter the two ratio-arm resistors and the adjustable standard resistor, and the calculator applies the balance condition.
The Wheatstone bridge is the most elegant idea in electrical measurement: do not measure the unknown at all — balance it against knowns until a meter reads nothing, and let a ratio do the work. Four resistors form a diamond; a supply drives one diagonal, a sensitive galvanometer watches the other. When the bridge balances, no current crosses the middle, and the unknown is fixed by the other three through Rx = R₂ × R₃ / R₁. Because the reading at balance is zero, the result depends on the accuracy of the standard resistors, not on the calibration of any meter — a trick so good that precision measurement used it for over a century, and strain gauges, RTDs and pressure sensors still report through bridge circuits today.
The wiring convention matters more than the algebra, so fix it now, exactly as this calculator uses it: R₁ and R₂ form one divider (top-left and bottom-left), R₃ and Rx form the other (top-right and bottom-right), and balance requires the two divider midpoints to sit at equal voltage. That condition, R₁/R₂ = R₃/Rx, rearranges to the calculator's formula. Swap any two arms in your head and you will solve for the reciprocal arrangement — the mistakes section shows how to catch that with a limiting-case check.
This page does three jobs. It gives the balance calculation instantly; it explains the sensitivity question — how far off balance a tiny change in Rx pushes the galvanometer, which is what makes bridges supreme for detecting changes (a strain gauge's 0.2% sigh becomes a readable signal); and it works the modern applications, from a Pt100 thermometer's 0.385 Ω-per-degree whisper to the four-gauge bridges inside every load cell and bathroom scale.
Formula
At balance, the ratios in the two arms are equal: R₁ / R₂ = R₃ / Rx, so Rx = R₂ × R₃ / R₁.
The balance condition comes from demanding equal voltages at the two midpoints: the left divider sets its midpoint at V × R₂/(R₁ + R₂), the right at V × Rx/(R₃ + Rx), and equating them gives the ratio. Note what is absent: the supply voltage and the galvanometer's properties appear nowhere in the result — they affect only how sharply the null can be found, not where it lies. Rearrangements solve for any arm from the other three: R₁ = R₂R₃/Rx, R₂ = R₁Rx/R₃, R₃ = R₁Rx/R₂. Off balance, the bridge output for small changes is approximately Vout ≈ Vs × (ΔR/R) ÷ 4 for a single active arm — the working equation of every strain-gauge instrument. Units: all resistances in ohms (or all in kΩ — ratios forgive a common unit, but never a mixed one).
| Symbol | Meaning | Units |
|---|---|---|
| R₁, R₂ | the ratio arms (one divider) | ohms (Ω) |
| R₃ | the known standard in the unknown's divider | ohms (Ω) |
| Rx | the unknown resistance | ohms (Ω) |
| ΔR/R | fractional change of an active arm (sensors) | dimensionless |
| GF | gauge factor of a strain gauge (≈ 2) | dimensionless |
Why a null beats a needle: sensitivity and the sensor bridge
A deflection meter's accuracy is capped by its own calibration — a 2% meter makes 2% measurements. A bridge at balance asks its detector only one question — "is it zero yet?" — which any sensitive galvanometer answers superbly, while the accuracy migrates to the standard resistors, which can be made to 0.01%. That division of labour is the bridge's genius, and it survives into the electronic age: modern instruments rarely balance to zero by hand, but they still use the bridge topology, reading the small off-balance voltage instead. For one active arm the output is Vs × (ΔR/R)/4: a 5 V excitation and a 0.2% resistance change give 5 × 0.002 ÷ 4 = 2.5 mV — small, but a clean, amplifiable signal sitting on a near-zero background, which is the easiest kind of signal in electronics to measure well.
Arrange more active arms and the output multiplies: two gauges in push-pull (one stretched, one compressed) double it, and a full bridge with four active arms gives Vout = Vs × (ΔR/R) — 10 mV for the same 0.2% change — while temperature effects, which move all four arms together, cancel in the ratio. This is why load cells, torque transducers and pressure sensors are built as full bridges: sensitivity, temperature compensation and a differential output in one four-resistor diamond, a hundred and eighty years after Wheatstone.
One refinement separates textbook from instrument practice: the linear rule Vout ≈ Vs(ΔR/R)/4 is an approximation. The exact single-arm expression is Vout/Vs = (ΔR/R) ÷ (4 + 2ΔR/R), so at a 10% resistance change the true output is about 5% below the linear prediction — irrelevant for a strain gauge's 0.2%, significant for a thermistor swinging tens of percent. Precision transmitters either keep changes small, use constant-current excitation tricks, or linearise in firmware; the bridge equation this calculator embodies is exact only at the null, and honest instruments remember it.
How to use
- Enter R₁ and R₂, the two resistors in the left-hand divider of the bridge.
- Enter R₃, the known resistor in series with the unknown.
- Click "Calculate" to get the unknown resistance Rx that balances the bridge.
- Sanity-check the result against the ratio: if R₂ is ten times R₁, Rx must come out ten times R₃. If it doesn't, two arms have been swapped — the ratio check takes five seconds and catches the classic bridge error.
Worked examples
Example 1 — the reference balance. A bridge balances with R₁ = 100 Ω, R₂ = 200 Ω and R₃ = 50 Ω. Find Rx.
Rx = R₂ × R₃ / R₁ = 200 × 50 / 100 = 100 Ω.
Check the ratios: R₁/R₂ = 100/200 = 0.5 and R₃/Rx = 50/100 = 0.5 ✓ — the two arms match, so the galvanometer reads zero.
Example 2 — the equal-arm bridge. With R₁ = R₂ = R₃ = 1,000 Ω: Rx = 1,000 × 1,000 ÷ 1,000 = 1,000 Ω. Equal ratio arms make Rx equal to the standard — the configuration sensor bridges are built in, because matched arms share temperature behaviour.
Example 3 — a strain gauge's whisper. A 350 Ω strain gauge with gauge factor 2 is strained by 1,000 microstrain (0.1%): ΔR/R = 2 × 0.001 = 0.002, so ΔR = 0.7 Ω — the gauge becomes 350.7 Ω. In a quarter bridge at 5 V excitation the output is 5 × 0.002 ÷ 4 = 2.5 mV; a full four-gauge bridge returns 10 mV. Weighing a truck with a 0.7 Ω change is the bridge's everyday miracle.
Example 4 — the Pt100 thermometer. A platinum RTD is 100 Ω at 0 °C and rises 0.385 Ω per degree. At 100 °C it reads 100 + 38.5 = 138.5 Ω. Placed in a bridge with R₁ = R₂, the standard arm that balances it is R₃ = Rx = 138.5 Ω — and a 1 °C change moves the arm by just 0.385 Ω, which is why RTD instruments use bridges (and 3- or 4-wire leads, below) rather than a plain ohmmeter.
Example 5 — a decade-box measurement. A precision bridge with ratio arms R₁ = 100 Ω, R₂ = 1,000 Ω (ratio ×10) balances at R₃ = 47.2 Ω. Rx = 1,000 × 47.2 ÷ 100 = 472 Ω — the ratio arms multiply the standard's reach, the same trick as a voltmeter's multiplier, in reverse.
Example 6 — a ×0.1 ratio. With R₁ = 1,000 Ω, R₂ = 100 Ω and balance at R₃ = 520 Ω: Rx = 100 × 520 ÷ 1,000 = 52 Ω. Small unknowns are measured with the ratio stepped down — and with extra care about lead resistance, which at these values is no longer negligible.
Wheatstone bridges in real life
Strain gauges and load cells. Every platform scale, crane weigher and structural test begins as Example 3: foil gauges bonded to a flexing element, wired as a full bridge, their 0.7 Ω whispers amplified into weights. The bridge's temperature cancellation is as valuable as its sensitivity.
RTD temperature measurement. Pt100 and Pt1000 sensors (Example 4) are read by bridges in process plants worldwide; the bridge rejects the supply variations and, in 3-wire form, the lead resistance that would otherwise swamp a 0.385 Ω/°C signal.
Pressure transducers. Piezoresistive dies are literally four resistors diffused into a diaphragm in a bridge layout — pressure flexes the diaphragm, two arms rise and two fall, and the full-bridge output is proportional to pressure.
Precision resistance metrology. The bridge's original job is not finished: standards laboratories still compare resistors by bridge methods when parts-per-million agreement is required, because no direct-reading ohmmeter touches a null comparison.
Fault location in cables. The Murray loop test is a Wheatstone bridge stretched along a faulty cable pair: the balance ratio locates an earth fault's distance from the test end — a 19th-century technique still in the fault-finder's kit.
Gas and humidity sensors. Pellistor gas sensors run a heated sensing bead against a compensator bead in a bridge; the gas burns on one bead only, unbalancing the bridge in proportion to concentration.
Lead resistance and the Kelvin answer
The bridge's ratio is exact; the wires connecting it are not. Every lead adds its own small resistance in series with an arm, and when the unknown is itself small — Example 6's 52 Ω, or a motor winding's fraction of an ohm — a metre of test lead can rival the quantity being measured. Classical bridges fought this with heavy, short connections and by folding lead resistance into the ratio arms where it cancels. The complete answer is Kelvin's: the four-wire (Kelvin) connection, in which current is driven through one pair of leads and the voltage is sensed through a separate pair that carries essentially no current, so lead resistance drops no sensed voltage. Modern micro-ohmmeters are Kelvin instruments, and 3-wire/4-wire RTD hookups are the same idea wearing bridge clothes. The lesson generalises beyond bridges: a measurement is a circuit, and everything in the circuit is part of the measurement.
From galvanometer to ADC: the bridge today
The swinging-coil galvanometer is gone from the bench, but the bridge it served is busier than ever — it simply reports to an analogue-to-digital converter now. Modern bridge transmitters excite the bridge with a precision reference and digitise the differential output ratiometrically: the same voltage that excites the bridge serves as the ADC's reference, so any drift in excitation cancels in the ratio — the digital reincarnation of the null method's indifference to supply voltage. Twenty-four-bit sigma-delta converters resolve the bridge's millivolts into millions of counts, finding the 0.7 Ω strain-gauge change of Example 3 with room to spare. Wheatstone's diamond survives because its virtues — ratio accuracy, differential signalling, temperature cancellation — are properties of the topology, not of the century.
One classical refinement survives the digitisation intact: reversing the excitation. Junctions of dissimilar metals in the wiring generate tiny thermal voltages that masquerade as bridge output; flipping the supply polarity flips the true signal but not the thermal offsets, so averaging the two readings cancels them. Precision transmitters still perform this chopper trick in firmware, for the same reason bridge operators once swapped their battery leads — the diamond rewards those who measure it twice, in opposite directions.
Quick reference: balance ratios
| Ratio R₂/R₁ | Rx in terms of R₃ | Typical use |
|---|---|---|
| 0.1 | Rx = 0.1 × R₃ | small unknowns (Example 6) |
| 1 | Rx = R₃ | sensors, equal-arm bridges |
| 10 | Rx = 10 × R₃ | decade-box reach (Example 5) |
| 100 | Rx = 100 × R₃ | large unknowns; watch R₃'s self-heating |
Extreme ratios trade convenience for accuracy: at ×100 the standard arm carries a very different current from the unknown, so self-heating drifts the two arms unequally. Precision practice keeps ratios between 0.1 and 10 where possible and lets R₃'s decade range do the rest.
Common mistakes
- Swapping the arms. The formula is Rx = R₂R₃/R₁ in this calculator's layout (R₁ over R₂ on the left, R₃ over Rx on the right). Wire or reason in a different layout and the answer inverts a ratio. Always run the limiting check: bigger R₂/R₁ must mean bigger Rx for the same R₃.
- Reading the galvanometer as the measurement. Off-balance deflection is not a calibrated resistance reading — the bridge measures at the null, or (in sensor service) through the calibrated approximation Vout ≈ Vs(ΔR/R)/4. Treating raw deflection as ohms mixes the two modes.
- Ignoring lead resistance on small unknowns. Two 0.2 Ω leads inside a 52 Ω measurement are a 0.8% error; inside a 0.5 Ω measurement they are the whole story. Use Kelvin connections or fold the leads into the ratio arms.
- Over-driving the bridge. A high supply voltage sharpens the null but heats the arms — and a heated standard is a changed standard. Sensor bridges also have a rated excitation (often 5–10 V) set by the gauges' self-heating, not by the electronics.
- Mixing units across the ratio. Ratios forgive ohms-versus-kΩ only if all four arms use the same unit. One arm entered in kΩ among ohms throws the result by a thousand, in whichever direction embarrasses most.
FAQ
What does "balanced" mean?
The bridge is balanced when the two middle points are at the same potential, so the galvanometer between them reads exactly zero. At that instant the balance equation holds precisely, whatever the supply voltage is.
What happens off balance?
If Rx ≠ R₃ × (R₂/R₁), the midpoints sit at different potentials and a current flows through the galvanometer — one way or the other depending on which side is larger. In the lab you adjust R₃ (or a ratio arm) until that current vanishes; the closer you get to zero, the more accurate the measurement.
Why is a bridge better than an ohmmeter?
It is a null method: the result depends only on the ratio of known resistors, not on the supply voltage or the galvanometer's calibration. That is why bridges can measure resistance far more accurately than a direct meter reading.
Do I have to use the exact arm positions shown?
The formula assumes R₁ and R₂ form one divider and R₃ and Rx the other, with the galvanometer between the midpoints. If your bridge is drawn differently, relabel it so the balance condition R₁/R₂ = R₃/Rx holds, or the formula will be wrong.
What is the bridge used for besides measuring resistance?
With one arm replaced by a sensor — a strain gauge, thermistor or RTD — the bridge converts tiny resistance changes into a voltage signal. Load cells, pressure sensors and precision thermometers are all Wheatstone bridges at heart.
What is a quarter, half and full bridge?
The count of active (changing) arms: one sensor with three fixed resistors is a quarter bridge; two sensors in push-pull make a half bridge; four active arms make a full bridge, with four times the quarter bridge's output and inherent temperature compensation.
How sensitive is a bridge to small changes?
For one active arm, Vout ≈ Vs × (ΔR/R)/4. At 5 V excitation a 0.01% change gives 12.5 µV — measurable with a decent amplifier, which is why bridges detect microstrain and millidegree temperature shifts.
Can a bridge measure capacitance or inductance?
Yes — AC bridges (Maxwell, Hay, Schering) balance impedances instead of resistances, with capacitors or inductors as the standards. The null philosophy is identical; the balance condition gains a second equation for the reactive part.
Why do RTD instruments use three wires?
To cancel lead resistance: two matched leads are placed in opposite arms so their resistances subtract out of the ratio, and the third carries the supply. It is the bridge compensating for its own wiring — see the lead-resistance section above.
Who invented it?
Samuel Hunter Christie described the diamond arrangement in 1833; Charles Wheatstone popularised it from 1843 and his name stuck. The same honour-by-publicity pattern gave the potentiometer circuit to Poggendorff — measurement history rewards its explainers.